6 - 20
Operator’s Manual
6 Image Acquisition
3.
Optimize the B mode image:
In the capture preparation status, select the B mode page tab to select B mode image
optimization. Do measurement or add comment/bodymark to the image if needed.
4.
Switch to [iScape] page to enter the iScape acquisition preparation status.
5.
Select [Start Capture] or [Update] button to begin the capture.
The system enters into image review status when the acquisition is completed. You can
perform operations such as parameter adjusting.
6.
Scan slowly to obtain a single extended field of view image. You can also erase and retrace if
the image is not satisfactory.
–
During image acquisition, none of the parameters are adjustable, and functions such as
measurement, comments and body marks are not available.
–
A green box on the image indicating the boundary between the merged images and the
unfinished images.
–
During image slicing, the system gives feedback on the probe’s moving speed in the form
of colors and words. The meanings are as follows:
7.
Do one of following to end image capture:
–
Select [Stop Capture].
–
Select [Update].
–
Wait until the acquisition completes automatically.
After the acquisition is completed, the panoramic image will be displayed and the system
enters iScape viewing mode.
6.9.2
Image Review
After the acquisition is completed, the panoramic image will be displayed and the system enters
iScape viewing mode.
In iScape viewing mode, you can perform image parameters setting, zooming, rotation and adding
measurements, comments and body marks.
Image size
Tap [Actual Size] to display the image in its actual size.
Tap [Fit Size] to display the image in a size that fits the current window.
Tint map
Tap [-] or [+] to change the value slightly or drag the control directly.
Ruler
Tap[Ruler] to hide or show the ruler around the image.
Rotation
Tap [-] or [+] to rotate the image clockwise or anticlockwise, or drag the control directly.
Status
ROI Color
Tip
Speed too low
Blue
Moving speed of the probe is too low.
Appropriate
Green
/
Speed too high
Red
Moving speed of the probe is too high.
Summary of Contents for TE X
Page 2: ......
Page 15: ...Contents Operator s Manual ix I Indications for use I 1...
Page 16: ......
Page 24: ...This page intentionally left blank...
Page 110: ...This page intentionally left blank...
Page 168: ...This page intentionally left blank...
Page 188: ...This page intentionally left blank...
Page 266: ...This page intentionally left blank...
Page 274: ...This page intentionally left blank...
Page 278: ...This page intentionally left blank...
Page 298: ...H 2 Operator s Manual H Probe Dimensions Length Height Max Width Max...
Page 328: ...This page intentionally left blank...
Page 329: ......
Page 330: ...P N 046 023006 00 2 0...