1 - 6
Operator’s Manual
1 Important Information
•
A transducer sheath must be installed over the transducer before
performing examination.
•
When using a probe, pay attention to the status of the ultrasound image.
Do not use the probe to perform image acquisition when the image is
frozen.
CAUTION
•
When using the probe, wear sterile gloves to prevent infection.
•
Please use the ultrasound gel com-pliant with the relevant local
regulations. And manage the ultrasound gel properly to ensure that it does
not become a source of infection.
•
In normal diagnostic ultrasound mode, there is no danger of a normal-
temperature burn; however, do not keep the probe on the same region of
the patient for more than 10 minutes to avoid risk of burn.
•
Do not use the carrying case for storing the transducer. If the carrying case
is used for storage, it may become a source of infection.
•
It is required to practice ALARA when operating ultrasound system.
Minimize the acoustic power without compromising the quality of images.
•
The probe and accessories supplied with it are not delivered disinfected or
sterilized. Sterilization (or high-level disinfect) before use is required.
•
Disposable components should be packaged sterile and for single-use
only. Do not use if integrity of packaging violated or if expiration date has
passed. Please use the disposable components compliant with the relevant
local regulations.
•
Please use the disinfection or sterilization solution recommended in this
operator's manual; otherwise Mindray will not be liable for damage caused
by other solutions. If you have any questions, please contact Mindray
Customer Service Department or sales representative.
•
Do not use pre-lubricated condoms as a sheath. Lubricant may not be
compatible with the probe material and damage may result.
•
The damage of the transducer may be caused by the contact of improper
gel or cleaner:
–
DO NOT dip the transducer in the strong polar solution of ethanol,
chloride of lime, ammonium chloride, acetone and formaldehyde.
–
DO NOT contact the transducer with solution or ultrasound gel
containing oily medium such as mineral oil or lanoline.
NOTE:
•
Read the following precautions to prevent the probe from malfunction:
–
Before connecting or disconnecting the probe, freeze or turn off the system.
–
Clean and disinfect the probe before and after each examination.
Summary of Contents for TE X
Page 2: ......
Page 15: ...Contents Operator s Manual ix I Indications for use I 1...
Page 16: ......
Page 24: ...This page intentionally left blank...
Page 110: ...This page intentionally left blank...
Page 168: ...This page intentionally left blank...
Page 188: ...This page intentionally left blank...
Page 266: ...This page intentionally left blank...
Page 274: ...This page intentionally left blank...
Page 278: ...This page intentionally left blank...
Page 298: ...H 2 Operator s Manual H Probe Dimensions Length Height Max Width Max...
Page 328: ...This page intentionally left blank...
Page 329: ......
Page 330: ...P N 046 023006 00 2 0...