Probes and Biopsy 13-29
NGB-007
Metal/needle detachable needle-guided bracket:
1. Put the sterile probe sheath on.
2. Hold the probe in one hand, select the correct needle-guided bracket and hold it with the other
hand. Match the groove and tab with the tab and groove of the probe respectively. Mount the
bracket onto the probe.
3. Screw the pinch nut of the needle-guided bracket to ensure that the needle-guided bracket is
properly installed on the probe.
4. Select a suitable guiding block and push it into the groove above the angle block, then clamp it
tightly.
5. Screw the block's nut to secure the block.
6. Insert a biopsy needle with the same specification as that of the guiding block into the guiding
block hole.
Plastic needle-guided bracket:
1. Put the sterile probe sheath on.
2. Hold the probe in one hand, select the correct needle-guided bracket and hold it with the other
hand. Align the tab at the narrow end of the needle-guided bracket with the groove of the
probe, then push the needle-guided bracket forward, so the tabs and grooves of the needle-
guided bracket align with the grooves and tabs of the probe.
3. Check manually to confirm that the needle-guided bracket is securely installed on the probe.
4. Select a suitable guiding block and push it into the groove above the angle block, then clamp it
tightly.
5. Insert a biopsy needle with the same specification as that of the guiding block into the guiding
block hole.
Summary of Contents for TE5
Page 1: ...TE7 TE5 Diagnostic Ultrasound System Operator s Manual Basic Volume ...
Page 2: ......
Page 6: ......
Page 12: ......
Page 24: ......
Page 36: ......
Page 54: ......
Page 110: ......
Page 115: ...Display Cine Review 6 5 6 Tap Return on the screen or tap Freeze to exit image compare ...
Page 120: ......
Page 124: ......
Page 156: ......
Page 174: ......
Page 192: ...12 18 Setup Select Advanced and do as follows Select MAPS and do as follows ...
Page 202: ...13 2 Probes and Biopsy C5 2s L12 4s L7 3s P4 2s L14 6s C11 3s L14 6Ns V11 3Ws P7 3Ts 7LT4s ...
Page 226: ...13 26 Probes and Biopsy NGB 034 NGB 035 ...
Page 250: ......
Page 272: ......
Page 276: ...A 4 Wireless LAN Tap Add Manually create a network profile to set ...
Page 282: ......
Page 318: ......
Page 322: ......
Page 323: ...P N 046 006959 07 1 0 ...