Electrical Safety Inspection E-3
E.5 Earth Leakage Test
Run an Earth Leakage test on the device being tested before performing any other leakage
tests.
The following outlet conditions apply when performing the Earth Leakage test.
normal polarity (Normal Condition);
reverse polarity (Normal Condition);
normal polarity with open neutral (Single Fault Condition);
reverse polarity with open neutral (Single Fault Condition).
LIMITS
For UL60601-1:
300 μA in Normal Condition.
1000 μA in Single Fault Condition.
For IEC60601-1:
500 μA in Normal Condition.
1000 μA in Single Fault Condition.
E.6 Enclosure Leakage Test
The following outlet conditions apply when performing the Enclosure Leakage test.
normal polarity (Normal Condition);
reverse polarity (Normal Condition);
normal polarity with open neutral (Single Fault Condition);
reverse polarity with open neutral (Single Fault Condition);
normal polarity with open earth (Single Fault Condition);
reverse polarity with open earth (Single Fault Condition).
LIMITS
For UL60601-1:
100 μA in Normal Condition.
300 μA in Single Fault Condition.
For IEC60601-1:
100 μA in Normal Condition.
500 μA in Single Fault Condition.
E.7 Patient Leakage Current
Patient leakage currents are measured between a selected applied part and mains earth. All
measurements have a true RMS only.
The following outlet conditions apply when performing the Patient Leakage Current test.
normal polarity (Normal Condition);
reverse polarity (Normal Condition);
normal polarity with open neutral (Single Fault Condition);
reverse polarity with open neutral (Single Fault Condition).
normal polarity with open earth (Single Fault Condition);
reverse polarity with open earth (Single Fault Condition).
Summary of Contents for TE5
Page 1: ...TE7 TE5 Diagnostic Ultrasound System Operator s Manual Basic Volume ...
Page 2: ......
Page 6: ......
Page 12: ......
Page 24: ......
Page 36: ......
Page 54: ......
Page 110: ......
Page 115: ...Display Cine Review 6 5 6 Tap Return on the screen or tap Freeze to exit image compare ...
Page 120: ......
Page 124: ......
Page 156: ......
Page 174: ......
Page 192: ...12 18 Setup Select Advanced and do as follows Select MAPS and do as follows ...
Page 202: ...13 2 Probes and Biopsy C5 2s L12 4s L7 3s P4 2s L14 6s C11 3s L14 6Ns V11 3Ws P7 3Ts 7LT4s ...
Page 226: ...13 26 Probes and Biopsy NGB 034 NGB 035 ...
Page 250: ......
Page 272: ......
Page 276: ...A 4 Wireless LAN Tap Add Manually create a network profile to set ...
Page 282: ......
Page 318: ......
Page 322: ......
Page 323: ...P N 046 006959 07 1 0 ...