Image Optimization 5-5
Operations
Drag the [Gain] control on the right part of the image area to adjust the gain.
Or, You can change gain value slightly by tapping on the gain control bar directly
Effects
Increasing the gain will brighten the image and you will see more received
signals. However, noise may also be increased.
Depth
Description
This function is used to adjust the sampling depth, the real-time value of which is
displayed in the image parameter area in the top-left corner of the screen.
Operations
Drag the [Depth] control on the right part of the image area to adjust the depth.
You can change depth value slightly by tapping on the depth control bar directly.
The adjustable depth values vary depending on the probe types.
Effects
Increase the depth to see tissue in deeper locations, or decrease the depth to
see tissue in shallower locations.
TGC
Description
The system compensates the signals from deeper tissue by segments to
optimize the image.
Tap
in the bottom-left corner of the image area to open the TGC
adjusting dialogue box.
There are 6-segment TGC sliders corresponding to the areas of the image.
Operations
To increase the gain compensation in an area of interest, drag the TGC control
to the right.
To decrease the gain compensation in the corresponding area of interest, drag
the control to the left.
Double-click any area on the dialogue box, all 6 TGC controls will return to
middle state.
About 1.5 seconds after the adjustment is complete, the TGC curve disappears.
Effects
Adjust the signal gain for a particular image area to achieve a balanced image.
Acoustic Power
Description
Refers to the power of ultrasonic waves transmitted by the probe, the real-time
value of which is displayed in the top-left part of the screen.
Operations
Adjust through [A.Power(%)] on the menu.
The adjusting range is 3.2-100.0.
Tap [-] or [+] to change the value slightly or drag the control directly.
Effects
Generally, increasing the acoustic power will increase the brightness and
contrast of the image and the force of penetration.
Impacts
You should perform exams according to actual situations and follow the ALARA
Principle.
Summary of Contents for TE5
Page 1: ...TE7 TE5 Diagnostic Ultrasound System Operator s Manual Basic Volume ...
Page 2: ......
Page 6: ......
Page 12: ......
Page 24: ......
Page 36: ......
Page 54: ......
Page 110: ......
Page 115: ...Display Cine Review 6 5 6 Tap Return on the screen or tap Freeze to exit image compare ...
Page 120: ......
Page 124: ......
Page 156: ......
Page 174: ......
Page 192: ...12 18 Setup Select Advanced and do as follows Select MAPS and do as follows ...
Page 202: ...13 2 Probes and Biopsy C5 2s L12 4s L7 3s P4 2s L14 6s C11 3s L14 6Ns V11 3Ws P7 3Ts 7LT4s ...
Page 226: ...13 26 Probes and Biopsy NGB 034 NGB 035 ...
Page 250: ......
Page 272: ......
Page 276: ...A 4 Wireless LAN Tap Add Manually create a network profile to set ...
Page 282: ......
Page 318: ......
Page 322: ......
Page 323: ...P N 046 006959 07 1 0 ...