5-8 Image Optimization
Line Density
Description
The function determines the quality and information of the image.
Operations
Adjust through [Line Density] on the menu.
There are four levels of line density available: UH, M, H, L.
Effects
The higher the line density, the higher the resolution.
Impacts
The higher the line density, the lower the frame rate.
Dynamic Range
Description
This function is used to adjust the B image resolution to compress or expand the
gray display range. The real-time value displays in the image parameter area in
the top-left corner of the screen.
Operations
Adjust through [Dyn Ra.] on the menu.
Tap [-] or [+] to change the value slightly or drag the control directly.
The adjusting range is 30-240, in increments of 5.
Impacts
The more the dynamic range, the more specific the information and the lower
the contrast with more noise.
iClear
Description
The function is used to enhance the image profile so as to distinguish the image
boundary for optimization.
Operations
Adjust through [iClear] on the menu.
Tap [-] or [+] to change the level slightly or drag the control directly.
The system provides 7 levels of iClear adjustment: off represents no iClear
effect, and the bigger the value the stronger the effect.
Impacts
It may cause increased noise.
Persistence
Description
This function is used to superimpose and average adjacent B images, so as to
optimize the image and remove noise.
Operations
Adjust through [Persistence] on the menu.
Tap [-] or [+] to change the level slightly or drag the control directly.
The system provides 7 levels of frame average adjustment: the bigger the value
the stronger the effect.
Effects
Persistence can remove image noise to make details clearer.
Impacts
Increasing Persistence may lead to missing signals.
TSI (Tissue Specific Imaging)
Description
The TSI function is used to optimize the image by selecting acoustic speed
according to tissue characteristics.
Operations
Select different TSI modes using the [TSI] control.
The system provides four ways of optimizing for specific tissues: general,
muscle, fluid and fat.
Summary of Contents for TE5
Page 1: ...TE7 TE5 Diagnostic Ultrasound System Operator s Manual Basic Volume ...
Page 2: ......
Page 6: ......
Page 12: ......
Page 24: ......
Page 36: ......
Page 54: ......
Page 110: ......
Page 115: ...Display Cine Review 6 5 6 Tap Return on the screen or tap Freeze to exit image compare ...
Page 120: ......
Page 124: ......
Page 156: ......
Page 174: ......
Page 192: ...12 18 Setup Select Advanced and do as follows Select MAPS and do as follows ...
Page 202: ...13 2 Probes and Biopsy C5 2s L12 4s L7 3s P4 2s L14 6s C11 3s L14 6Ns V11 3Ws P7 3Ts 7LT4s ...
Page 226: ...13 26 Probes and Biopsy NGB 034 NGB 035 ...
Page 250: ......
Page 272: ......
Page 276: ...A 4 Wireless LAN Tap Add Manually create a network profile to set ...
Page 282: ......
Page 318: ......
Page 322: ......
Page 323: ...P N 046 006959 07 1 0 ...