Trio™ Operating Instructions
0070-10-0666-01
6 - 27
Appendix
Physical Specifications
Isolation:
Risk (Leakage) Currents:
Meets the requirements of IEC 60601-1/
EN 60601-1
Enclosure Risk Current:
Normal operating conditions
<= 100
μ
A
Single fault condition
<= 300
μ
A for 120 VAC
Single fault condition
<= 500
μ
A for 230 VAC
Patient Source Current:
Normal operating conditions
<= 10
μ
A
Single fault condition
<= 50
μ
A
Patient Sink Current:
Normal operating conditions
not applicable
Single fault condition
<= 50
μ
A
Dielectric Withstand
Per IEC 60601-1/EN 60601-1