Trio™ Operating Instructions
0070-10-0666-01
2 - 9
Operations
System Menu
NOTE:
The recorder is optional.
Select
PREVIOUS MENU
to return to the previous menu. Select
NORMAL SCREEN
(from
the menu or the Front Panel Keypad) to exit the menu and return to the normal screen.
TIME
TIME
represents the length of recording time. There are two (2)
selections available:
CONTINUAL
and
8s
.
CONTINUAL
means
once the user presses the
key on the front panel, the
recorder will continuously print out the selected waveform(s) until the
button is pressed again.
8s
indicates a waveform recording time
length of 8 seconds.
INTERVAL
INTERVAL
represents the time interval between the start of recorder
activations. The following 10 selections are available:
OFF,
10min, 20min, 30min, 40min, 50min, 1HR, 2HRS, 3HRS
and
4HRS
. The system will initiate the print operation according to
the selected time interval. All interval waveform printouts are 8
seconds in length.
NOTE:
A real-time printout (based on TIME
as described in the previous menu)
takes priority over a printout based
on INTERVAL.
SPEED
SPEED
is used to select the speed at which the paper advances in
the printer. There are two (2) options:
25.0
and
50.0
mm/s
.
GRID
GRID
is used to select output format:
OFF
produces a printout
without a background grid, and
ON
produces a printout with a
background grid.
CLEAR PRINT
TASK
CLEAR PRINT TASK
is used to clear all print tasks from the
recorder. This can be used, for example, when multiple alarm print
tasks are triggered simultaneously.