Trio™ Operating Instructions
0070-10-0666-01
2 - 65
Operations
Alarms
COMMON ALARM SETUP Menu Selections
FIGURE 2-47
Common Alarm Setup Menu
ALM SEL
(Alarm Selection)
Choose
COMMON ALARM SETUP
.
ALARM VOL
Volume of the audio alert. Selections are:
LOW
,
MED
, and
HIGH
.
ALM PRINT TIME
Length of the printed waveform. Selections are:
8s
,
16s
, or
32s
.
ALM MUTE TIME
Length of time for the alarm to be silenced. Selections are:
1min
,
2min
,
5min
,
10min
, and
PERMANENT
.
When
ALM MUTE TIME
is set to
PERMANENT
and Alarm Mute
is active, the message “Alarms Permanently Muted” displays at a
constant rate in the technical alarms area.
• If another technical alarm occurs at the same time, both
messages will display alternately.
• If the
ALM MUTE TIME
is changed to something other than
PERMANENT
, the following conditions will exist:
• The elapsed time is defined as the time from which
ALM
MUTE TIME
had been set to
PERMANENT
to the time at
which it was changed to something other than
PERMANENT
.
• If the elapsed time exceeds the newly selected
ALM MUTE
TIME
, then the Alarm Mute will immediately deactivate.
Example: Elapsed time = 5min, New
ALM MUTE TIME
=
2min, Alarm Mute immediately deactivates.
• If the elapsed time is less than the newly selected
ALM
MUTE TIME
, then the Alarm Mute will continue to be active
for the remainder of the newly selected time period.
Example: Elapsed time = 2min, New
ALM MUTE TIME
=
10min, Alarm Mute continues to be active for 8 minutes and
then deactivates.