Trio™ Operating Instructions
0070-10-0666-01
2 - 69
Operations
Alarms
NIBP ALARM SETUP Menu Selections
FIGURE 2-51
NIBP Alarm Setup Menu
ALM SEL
(Alarm Selection)
Choose
NIBP ALARM SETUP
.
ALM
Allows the user to turn the alarm
ON
or
OFF
. Choose
ON
to
enable the alarm; choose
OFF
to disable the alarm. If the alarm is
set to
OFF
, the alarm
OFF
symbol
will display within the
parameter tile. The alarm is activated when the measurement
exceeds the high or low limits.
ALM PRIORITY
Allows the user to select the priority of the alarm. Choices are
1
,
2
and
3
. Priority 1 alarms are considered the most serious.
ALM PRINT
Allows the user to enable or disable automatic printing during a
physiological alarm condition. Choose
ON
to enable printing
upon alarm. Choose
OFF
to disable printing upon alarm. See
“Recorder (Optional)” on page 2-82 for details.
SYS ALM HI
Allows the user to set the upper limit of the Systolic
NIBP
alarm.
(See “NIBP Alarm Limits” on page 2-50.)
SYS ALM LO
Allows the user to set the lower limit of the Systolic
NIBP
alarm.
(See “NIBP Alarm Limits” on page 2-50.)
MEAN ALM HI
Allows the user to set the upper limit of the Mean
NIBP
alarm.