3-4
3.2.2 Interlock Device
Ohmeda Selectatec
®
compatible interlock device
is used. When the anesthetic delivery system is
connected to multiple vaporizers, if one vaporizer is switched on, the two pins on the interlock device
are pushed out, preventing other vaporizers from being switched on.
WARNING
z
Before operation, check if the interlock device is fully functional..
z
A malfunction in the interlock system can endanger the patient by overdosing or
mixing anesthetic agents.
NOTE
z
To ensure the normal operation of the vaporizer, connect the vaporizer in the correct
flow direction corresponding with the arrows on the Anesthetic Vaporizer.
z
When the anesthetic delivery system has three and more than three groups of plug-in
connectors, check if there is an interlock function between nonadjacent connectors.
Otherwise, vaporizers are recommended to be connected right next to each other.