6-7
NOTE
z
If there are no valves on the anesthesia machine plug-in connectors, fresh gas and
anesthetic agent vapor may escape when the vaporizer is removed.
6.5 Moving when Filled
This operation is only to be done as part of normal operation, not for storage and transport.
1.
The anesthetic delivery system can be moved at the workplace with the vaporizer switched
on.
NOTE
z
Jerky movements or tilting at an angle of more than 30
°
can cause incorrect output
concentration.
2. The anesthetic delivery system with securely fastened vaporizers can be moved with control
dial set at “0”, if there is no risk of tilting by more than 30
°
.
WARNING
z
When tilted at an angle of more than 30
°
:
The anesthetic agent may overflow when the control dial is set at “0”, endangering
health.
When the control dial is set above “0”, the anesthetic agent may leak and get into the
flow control system and cause excessively high or low concentrations when the
vaporizer is used next time.
3. When the vaporizer is detached from the anesthetic delivery system and transported
separately, the control dial must remain at the “0” position.
Summary of Contents for V60
Page 1: ...V60 Anesthetic Vaporizer Operator s Manual ...
Page 2: ......
Page 9: ...3 C Symbols and Terminology C 1 C 1 Symbols C 1 C 2 Terminology C 1 ...
Page 10: ...4 FOR YOUR NOTES ...
Page 17: ...2 3 2 3 2 Rear View 1 Handle for locking lever 2 Interlock system 3 Locking pin ...
Page 48: ...5 4 FOR YOUR NOTES ...
Page 56: ...6 8 FOR YOUR NOTES ...
Page 66: ...9 4 FOR YOUR NOTES ...
Page 79: ...11 11 0 1 2 3 4 5 6 0 1 2 3 4 5 6 G C B A F Concentration delivered vol Enflurane Time h ...
Page 80: ...11 12 ...
Page 84: ...A 4 FOR YOUR NOTES ...
Page 86: ...B 2 FOR YOUR NOTES ...
Page 89: ......
Page 90: ...P N 046 001002 00 8 0 ...