8-3
2. Measuring
(1)
Check the output concentration at “0” position, 0.4, 1, 2, 3, 5, and MAX in ascending
order.
(2)
Correct measured values, according to the carrier gas..
Air check: no correction.
O
2
check: reduce the measured values as follows:
Measured value vol%
Correction
<1%
1.0-2.0
2.0-4.0
5.0-8.0
-0.05
-0.10
-0.20
-0.30
If the data displayed is in % partial pressure, no correction is made. If it is in vol.%, it needs to be
converted to partial pressure. The formula is:
Concentration (% partial pressure) =
Measured value (vol.%) x atmospheric pressure (kPa)
101.3 kPa
3. Determine the accuracy range.
Range of concentration accuracy (maximum value always applies)
Operating environment
15 to 35
℃
or 0.2 to 10 L/min
10 to 15
℃
or 35 to 40
℃
or 10 to
15 L/min
Set concentration
≤
6%
±0.20 vol.% or ±20% rel.,
whichever is greater
+0.30/-0.20 vol.% or +25/-20%
rel., whichever is greater
Set concentration > 6%
±0.25 vol.% or ±20% rel.,
whichever is greater
+0.35/-0.25 vol.% or +30/-20%
rel., whichever is greater
4. Test
result
If the corrected measured value is within the permissible range of output concentration, the
vaporizer can be put into operation.
CAUTION
z
If the corrected measured value is not within the permissible range of output
concentration, do not use the vaporizer. Have the vaporizer checked by trained service
personnel.
Summary of Contents for V60
Page 1: ...V60 Anesthetic Vaporizer Operator s Manual ...
Page 2: ......
Page 9: ...3 C Symbols and Terminology C 1 C 1 Symbols C 1 C 2 Terminology C 1 ...
Page 10: ...4 FOR YOUR NOTES ...
Page 17: ...2 3 2 3 2 Rear View 1 Handle for locking lever 2 Interlock system 3 Locking pin ...
Page 48: ...5 4 FOR YOUR NOTES ...
Page 56: ...6 8 FOR YOUR NOTES ...
Page 66: ...9 4 FOR YOUR NOTES ...
Page 79: ...11 11 0 1 2 3 4 5 6 0 1 2 3 4 5 6 G C B A F Concentration delivered vol Enflurane Time h ...
Page 80: ...11 12 ...
Page 84: ...A 4 FOR YOUR NOTES ...
Page 86: ...B 2 FOR YOUR NOTES ...
Page 89: ......
Page 90: ...P N 046 001002 00 8 0 ...