If the applied load exceeds the minimum load limit:
-
the weighing platform is automatically tared, provided the [AUT.TARA] menu item
(3.7.2) is activated (see Chapter
-
an automatic protocol printout is generated, provided the menu item [AUT.EINM]
(7.15.2) is activated (see Chapter
7.1.3
Automatic printout
Menu: [SETUP]- [DRUCK]- [PROTOK.]- [AUT.EINM] (7.15) (see also Chapter
If the menu item [EIN] (7.15.2) is active, the irst weight value exceeding the minimum
load is printed. If the menu item for automatic taring is also activated, only taring is
performed when the minimum load is exceeded.
In this case, an automatic printout is not generated until the second weight value that
exceeds the minimum load.
7.1.4
Adjustment/Coniguration Counter for Standard Scales
Automatic recording of changes to adjustment and weighing parameters using two
independent counters. The values remain saved for the life of the component.
Adjustment Counter Features:
-
Counter limited to 9999
-
Counter at "C 0000" for hardware commissioning
-
Counter cannot be reset
-
Counter is updated automatically when:
Menu: [SETUP] - [WP-1] - [INTERN.] - [PARAM.1] (1) (see chapter
-
linearization, calibration/adjustment is successful
-
user calibration, adjustment or linearization weight is changed (menu item
[MAN.EXT.W] 1.18)
-
When the following parameters are changed:
Zero setting range (menu item [ZERORNG] 1.11)
Tare/zero at power on (menu item [INIT.ZER] 1.12)
The above parameters are reset to factory settings (menu item [RESET] -
[WT.PARA] - [YES] 9.1.1) (see chapter
).
Coniguration Counter Features:
-
Counter limited to 9999
-
Counter at "P 0000" for hardware commissioning
-
Counter cannot be reset
-
Counter is updated automatically when:
-
Whenthefollowingparametersarechanged:
Menu: [SETUP] - [WP-1] - [INTERN.] - [PARAM.1] (1) (see chapter
)
Installation location (menu item [AMBIENT] 1.1)
Application ilter (menu item [ APP.FILT] 1.2)
Stability range (menu item [STAB.RNG.] 1.3)
Combics 1 Series Complete Scales Models CAW1P, CAW1S, CAS1
7 Operating
EN-84
Minebea Intec