7.2 Calibration, adjustment
During
calibration
, a possible deviation between the displayed measured value and the
true weight of a sample (calibration/adjustment weight) becomes evident. Calibration
does not entail making any changes to the scale.
During
adjustment
, a deviation between the displayed measured value and the true
weight of a sample (calibration/adjustment weight) is corrected or reduced to permissible
limit values.
or
Display weight value in second weight unit (here, e.g.kg)
or
Display of weight value with 10x resolution
This display switches back automatically after 10 seconds.
7. Press the
key to switch to the previous display.
Display of the net weight value as before switching.
8. To print a protocol, press the
key.