Characteristics
The characteristics available are speciied by the connected weighing platform and can
each be set in the menu [SETUP]:
-
No external adjustment possible with veriied scales
-
External adjustment with the standard weight of the factory setting or a weight
speciied by the user (not for veriied scales):
Menu [SETUP]- [WP-1]- [INTERN.]- [PARAM.1]- [CAL.JUST] (1.9) "Calibration,
adjustment" (see also Chapter
-
Weight speciication for external calibration/adjustment:
Menu [SETUP]- [WP-1]- [INTERN.]- [PARAM.1]- [HND.EXT.G] (1.18) "Manual input of
external weight" (see also Chapter
-
Internal calibration for IS weighing platforms (default setting under: [COM1] menu
(see Chapter
)
-
Lock the
key so that the functions speciied above cannot be triggered:
Menu [SETUP]- [WP-1]- [INTERN.]- [PARAM.1]- [CAL.JUST]- [GESPRT] (1.9.10) "Key
locked" (see also Chapter
-
Calibration and automatic or manual adjustment (not for veriied scales):
Menu [SETUP]- [WP-1]- [INTERN.]- [PARAM.1]- [CAL.ABL.] (1.10) "Calibration/
adjustment procedure" (see also Chapter
).
-
Release or block external adjustment:
Menu [SETUP]- [WP-1]- [INTERN.]- [PARAM.1]- [CAL.EXT] (1.16) "Enable external
calibration" (see also Chapter
Example:
External calibration (see Chapter
7.2.1
Set "Mandatory calibration" function
There is a switch to change over to "Goods weighing in legal metrology" . This is located
under a cover cap to the rear left of the weighing platform housing.
Use of the veriied scale in legal metrology in the EU:
The type approval for conformity assessment is valid only for non-automatic scales; for
automatic operation with or without additional attached devices, the national regulations
applicable to the place of installation must be observed.
Externally connected IS scales:
Before you use the scale in legal metrology, use the internal calibration equipment to
calibrate the scale at the site of installation (see Chapter
).
NOTICE
Damage due to increased temperature range
The temperature range indicated on the identiication plate (°C) must not be exceeded
during operation.
Monitor the temperature during operation and cool the installation site if necessary.
7 Operating
Combics 1 Series Complete Scales Models CAW1P, CAW1S, CAS1
Minebea Intec
EN-89