7.3 Internal Calibration/Adjustment for CAS… Models
System Requirements
-
"Internal Calibration" must be set in the Setup:
Menu [SETUP] - [WP-1] - [INTERN.] - [PARAM.1] - [CAL./ADJ] - [CAL.INT.] (1.9.4) (see
chapter
).
-
The scale housing has a built-in motorized calibration weight.
Example:
Internal calibration and then adjustment (see chapter
).
7.3.1
Example: Internal calibration
Internal calibration and then adjustment (weighing parameters: factory settings for CAS…
models)
(e.g., when adjustment prompt lashes WP). [CAL.INT.] is displayed for two seconds.
--------------------
Typ CAW1P1-3DC-L
Ser.no. 12345678
Vers. C1 111.031112
BVers. 01-62-02
Ser.no. A 12345678
--------------------
Ext. Kalibrieren
Soll + 10000 g
Diff. + 1 g
Ext. Justieren
Diff. + 0 g
--------------------
Name:
--------------------
1. Waage nullstellen with
key.
2. Startcalibration with
key.
The internal calibration weight is loaded automatically.
Combics 1 Series Complete Scales Models CAW1P, CAW1S, CAS1
7 Operating
EN-92
Minebea Intec