7.2 Calibration, adjustment
During
calibration
, a possible deviation between the displayed measured value and the
true weight of a sample (calibration/adjustment weight) becomes evident. Calibration
does not entail making any changes to the scale.
During
adjustment
, a deviation between the displayed measured value and the true
weight of a sample (calibration/adjustment weight) is corrected or reduced to permissible
limit values.
Characteristics
The characteristics available are speciied by the connected weighing platform and can
each be set in the menu [SETUP]:
-
No external adjustment possible with veriied scales
-
External adjustment with the standard weight of the factory setting or a weight
speciied by the user (not for veriied scales):
Menu [SETUP]- [WP-1]- [INTERN.]- [PARAM.1]- [CAL.JUST] (1.9) "Calibration,
adjustment" (see also Chapter
-
Weight speciication for external calibration/adjustment:
Menu [SETUP]- [WP-1]- [INTERN.]- [PARAM.1]- [HND.EXT.G] (1.18) "Manual input of
external weight" (see also Chapter
-
Internal calibration for IS weighing platforms (default setting under: Menu [COM1]
(see Chapter
) or [UNICOM]- [WP-2]
7. To delete the tare weight, enter
using the number block and press
key.