9 Maintenance/repairs/cleaning
9.1 Maintenance
Regular maintenance preformed on your device by an employee of Minebea Intec
Customer Service will ensure continued measurement accuracy. Minebea Intec can
provide you with maintenance contracts with cycles from one month up to two years. The
frequency of the maintenance intervals depends on the operating conditions and the
tolerance requirements of the operator.
9.2 Repairs
Maintenance work may only be carried out by specialists:
-
who have access to the necessary maintenance documents and instructions and
-
who have attended appropriate training sessions.
Note:
The seals placed on the device indicate that the device may only be opened and serviced
by authorized specialists to ensure trouble-free and safe operation of the device, and to
ensure that the warranty remains valid.
WARNING
Improper repairs can pose considerable risks to the user.
Immediately disconnect a defective device from the power supply (disconnect the
power plug from the electrical outlet).
Repairs should only be performed by Minebea Intec authorized technical personnel
using original spare parts.
Defective or damaged cables or screw connections must be replaced as a complete
unit.
WARNING
Working on a device that is switched on can have life-threatening consequences
Do not open the device while under power.
Wait at least 10 seconds after disconnecting from the power supply before opening.
The device must be properly opened and closed because the itting surfaces on the
housing parts inluence the IP protection.
9.2.1
Changing the protective cover
A damaged protective cover should be replaced immediately.
.
Remove the damaged protective cover.
.
Press the new protective cover over the edge at the front and rear of the until it is
irmly in position.
Combics 2 Series Complete Scales Models CAW2P, CAW2S
9 Maintenance/repairs/cleaning
EN-192
Minebea Intec