the point at which the luctuation of a measured value lies within a deined tolerance. The
narrower the tolerance, the more stable the platform is at "stability".
The [ACC.STAB] (3.11.2) "increased stability" setting has a lower tolerance so that the
average piece weight saves is more accurate and the results more reproducible; however,
the measurement time can take longer.
Setting: [APPLIC.] - [APPLIC.1] - [COUNT.] - [SAVE WT] (3.11) (see chapter
or
7.7.5.4
Reference Sample Updating
You can deine whether or not the reference sample weight is updated automatically
during weighing using this setting [REF.UPDT] (3.12). The reference sample weight is
updated automatically only when the following criteria are met:
-
Reference sample updating [AUTOMAT] (3.12.2) "Automatic" must be set for
reference sample updating in the menu.
-
The current piece count exceeds the original piece count by at least two.
-
The current piece count cannot be more than double the size of the original piece
count. This limitation does not apply to the irst update if the piece weight was
entered via a barcode scanner or keypad.
-
The current piece count is less than 1000 pcs.
-
The internally calculated piece count (such as 17.24 pcs) difers by less than ± 0.3 pcs
from the nearest whole number (see chapter
).
-
The weighing platform is stable in accordance with the parameter deined for saving
weights.
If automatic reference sample updating is selected in the menu and the piece count (pcs)
is displayed, the [AUTO] symbol is displayed below the bar graph. If the average piece
weight has been updated since you began weighing, the text line shows the [0PT]
symbol. During an updating operation, [0PT] and the updated piece count are displayed
briely in the measured value line. The new reference sample weight and reference
sample quantity are saved.
Setting: [APPLIC.] - [APPLIC.1] - [COUNT.] - [REF.UPDT] (3.12) (see chapter
or
7.7.5.5
Counting with Two Weighing Platforms
You can use two weighing platforms simultaneously with the Counting application. When
using two platforms, you can choose from the following operating modes:
-
Counting with two platforms of the same type (see chapter
-
Counting with one reference platform and one weighing platform (see
chapter
Note:
If automatic reference sample updating is enabled, the update is performed on the
active platform; in other words, the system does not automatically switch to the
reference platform.
7.7.5.5.1 Counting with two platforms of the same type
Use this mode to count diferent types of sample material with diferent weights. For
example, count the lighter-weight pieces on one platform and the heavier pieces on
another. You can deine one of the two scales as the reference scale.
7 Operating
Combics 2 Series Complete Scales Models CAW2P, CAW2S
Minebea Intec