The reference scale is the irst scale active when you switch on the device, regardless of
the setting for automatic initialization of the application.
7.7.5.5.2 Counting with one reference platform and one weighing platform
In this operating mode, the reference platform is a high-resolution weighing platform
with a relatively low maximum capacity. The other platform is used for weighing heavier
samples, and has a high capacity with a relatively low resolution. This allows you to both
determine the reference sample weight with high resolution; i.e., very precisely, and to
count large amounts of parts, without requiring an expensive high-resolution, high-
capacity weighing platform.
The system can be conigured to switch automatically to the reference platform for
initialization (the measured value line shows [REF]). Following initialization, you can
switch to the counting platform.
Setting: [APPLIC.] - [APPLIC.1] - [COUNT.] - [REF.WP] (3.13) (see chapter
oder
).
7.7.6
Example: Counting
You need to determine an unknown number of parts and the measurements should be
logged.
Coniguration
: The "Counting" application is selected, and printout has been set up (see
chapter
).
This is not required if the automatic tare function is active. The tare weight is saved
automatically when you place the container on the platform.
1. Place empty container on the scale.
2. Tare the scale with
key.
3. Place a number of parts in the container for the reference quantity (in this example,
20 pcs).