— 2 DIGIT
3.6.2 - 2 digits
— 5 DIGIT
3.6.3 - 5 digits
— 10 DIGT
3.6.4 - 10 digits
— 20 DIGT
3.6.5 - 20 digits
— 50 DIGT
3.6.6 - 50 digits
— 100 DIG
3.6.7 - 100 digits
— 200 DIG
3.6.8 - 200 digits
— 500 DIG
3.6.9 - 500 digits
— 1000 DI
3.6.10 - 1000 digits
— START
3.18 - Start averaging
— MANUAL
3.18.1 - Manual*
— AUTOMAT
3.18.2 - Automatic
— ACTIVTY
3.19 - Animal activity
— 0.1 PERC
3.19.1 - 0,1% of the animal/object
— 0.2 PERC
3.19.2 - 0,2% of the animal/object*
— 0.5 PERC
3.19.3 - 0,5% of the animal/object
— 1 PERC.
3.19.4 - 1% of the animal/object
— 2 PERC.
3.19.5 - 2% of the animal/object
— 5 PERC.
3.19.6 - 5% of the animal/object
— 10 PERC
3.19.7 - 10% of the animal/object
— 20 PERC
3.19.8 - 20% of the animal/object
— 50 PERC
3.19.9 - 50% of the animal/object
— 100PERC
3.19.10 - 100% of the animal/object
3.20 - Autom. printout of results
— MANUAL
3.18.1 - Manual*
— AUTOMAT
3.18.2 - Automatic
— DIS.UNLD
3.21 - Static display of result after load removed
— CLEARED
3.21.1 - Display is ixed until unload threshold
reached*
— PRESENT
3.21.2 - Fixed display until
is pressed
* Factory setting
7.9.6.1
Minimum load for application
The menu item [MIN.INIT] (3.6) sets the minimum load, i.e. the load that must be placed
on the weighing platform in order to carry out the application.
If the load on the platform is too light, the following will occur:
-
Error code [INF 29] appears,
-
The weighing platform is not initialized,
-
The preset reference sample quantity is saved.
Setting: [APPLIC.] - [APPLIC.1] - [ANIM.WG] - [MIN.INIT] (3.6) (see chapter
or
).
The minimum load can be set in 10 steps from 1 to 1000 digits (see available parameters).
The "digits" here refer to the scale intervals for the connected weighing platform. If the
interval of the connected platform is 1 g, for example, and 1000 digits are required, you
1. Select menu [APPLIC.] - [APPLIC.1] - [ANIM.WG] with
key and open with
key (see chapter
2. To save the setting, press the
key.
3. To exit setup: Press the
key several times.
7 Operating
Combics 2 Series Complete Scales Models CAW2P, CAW2S
Minebea Intec
EN-127