must place at least 1000 g (= 1000 intervals = 1000 digits) on the weighing platform for
initialization.
7.9.6.2
Starting the Measurements
The averaging routine does not begin until the luctuation in weight value remains below
a deined threshold over three consecutive measurements. The tolerance limit is deined
as a percentage of the animal or object weight (0,1 %; 0,2 %; …; 50 %; 100 %)
Setting: [APPLIC.] - [APPLIC.1] - [ANIM.WG] - [ACTIVTY] (3.19) (see chapter
or
If the "Averaging" parameter is set to 2 %, for example, and the animal or object weighs
10 kg, measurement does not begin until the luctuation in weight value remains below
200 g during three consecutive measurements.
7.9.6.3
Display
A calculated average value is shown continuously on the main display. The
symbol
indicates the calculated value.
You can toggle between the results display and the current scale display by pressing th
key.
Setting: [APPLIC.] - [APPLIC.1] - [ANIM.WG] - [DIS.UNLD] (3.21) (see chapter
or
).
You can select [CLEARED] (3.21.1) "Display is static until unload threshold reached" to
have the display switch automatically to the weight readout when you unload the
weighing platform (i.e., when the load is less than half the minimum load). The result of
the most recent averaging operation is not saved.
f you select [PRESENT] (3.21.2) "Display is static until the
key is pressed," he
calculated average remains displayed even after the weighing platform is unloaded, until
you press the
key to begin a new measurement.
7.9.7
Example: Averaging (Animal Weighing)
The weight of one mouse should be measured.
Coniguration
: The "Averaging" (Animal Weighing) application is selected, and printout
has been set up (see chapter
This is not required if the automatic tare function is active. The tare weight is saved
automatically when you place the container on the platform.
1. Place empty container on the scale.
2. Tare the scale with
key.
Combics 2 Series Complete Scales Models CAW2P, CAW2S
7 Operating
EN-128
Minebea Intec