8 Data interfaces
8.1 General information
The is equipped with the following data interfaces:
- COM1
: Standard data interface (RS
‑
232)
- UniCOM
: Universal data interface (optional)
Both interfaces can be conigured in the [SETUP] menu for diferent input and output
functions (e.g. printer, 2nd weighing platform, PC, control display). The optional UniCOM
interface can be used as RS
‑
232, RS
‑
485 or RS
‑
422, analog output (voltage/current
interface), galvanically isolated digital I/O, Proibus, DeviceNet or Ethernet.
A bar code reader can be connected via the PS/2 port or via the corresponding screw
terminals (IP69K).
Characteristics
-
CAW*P s (IP44 protection):
Connection via 25
‑
pin D
‑
sub socket.
-
CAW*S (IP69K protection):
The connection cable of the peripheral device is inserted into the via a screw
connection. The free cable ends are connected via screw terminals.
When using third-party or commercially available RS
‑
232 connection cables:
NOTICE
The pin assignments are often not compatible with Minebea Intec devices.
Non-compliance can lead to errors or to the damage to the and any peripheral devices
connected to it.
The pin assignment should be checked against the connection diagrams and
unassigned cables should be disconnected.
8.2 Speciication
8.2.1
Serial interface
Interface operation:
full duplex
Levels:
COM1:
RS
‑
232
UniCOM
1)
: RS
‑
232 or RS
‑
422/485
Transmission rate:
150, 300, 600, 1200, 2400, 4800, 9600, 19,200 baud
(depending on operating mode)
Number of databits:
7/8 bit
Parity:
Space, Odd, Even, None (depending on operating mo-
de)
Number of stop bits:
1 or 2 stop bits
Combics 2 Series Complete Scales Models CAW2P, CAW2S
8 Data interfaces
EN-172