Output of a conigurable printout protocol with or without standstill (menu item
[MAN./AUT]- [PROT.DRK] 6.1.7) and 6.1.8). Output is linked to the menu item "Print
protocols" ([DAT.PROT]) (see Chapter
).
With the exception of the output of a conigurable print protocol, the current display
value (weight value with unit, calculated value, alphanumeric display) is output.
8.3.2
SMA communication
Standardized communication protocol of the
S
cale
M
anufacturers
A
ssociation.
8.3.3
Proibus
See special description.
8.3.4
Ethernet
With the SBI, SMA and XBPI communications protocol
Modbus/TCP similar to the proibus data format
8.3.5
DiviceNet
See MAN-YDO…-C…_YDO…SW…
8.4 Data input format
A computer connected via the data interface (SBI communication) can send control
commands to the in order to control the weighing or application functions.
All commands have a shared frame format (data input format). They start with the
character
ESC
(ASCII: 27) and end with the character sequence
CR
(ASCII: 13) and
LF
(ASCII: 10). Their length varies from min. 4 characters (1 command character) up to max.
7 characters (4 command characters). When sending texts, this igure may also be higher.
The commands shown in the following table must each have the frame format ESC … CR
LF added.
Example
: The output command "P" ("Send display value") must be sent to the . The
character sequence "ESC P CR LF" is sent for this purpose.
Control commands
Command
Meaning
K
Weighing mode 1
L
Weighing mode 2
M
Weighing mode 3
N
Weighing mode 4
O
Lock keyboard
P
Send display value to the data interface
Q
Output acoustic signal
R
Release keyboard
T
Tare and zero (tare combi-function)
f3_
Zero, like command "kZE_"
f4_
Tare (without zeroing), like command "kT_"
Combics 2 Series Complete Scales Models CAW2P, CAW2S