Example
:
Output of the weight value +1255.7 g
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22
G # * * * * + * * * 1 2 5 5 . 7 * g * * CR LF
Position 1 … 6: Marking, right-aligned illed with blank spaces
Position 7: Algebraic sign +, – or blank space
Position 8: Blank space
Position 9 … 16: Weight value with decimal point. Leading zeros are displayed as blank
spaces (instead of a decimal point, a comma can be used, see Chapter
menu item
[DEZIM.TZ] 7.17).
Position 17: Blank space
Position 18 … 20: Character for measurement unit, blank space or character
!
as symbol
Position 21: Carriage Return
Position 22: Line Feed
Note:
If the weight value is displayed with a resolution 10
‑
times higher, it may not be printed
or stored on scales used in legal metrology that are operated in SBI mode. In this case,
the unit character is never output with the data output.
8.5.3
External Keyboard Functions (PC Keyboard)
Setting: [SETUP] - [BARCODE] - [EXT.KEYB] (see chapter
)
The alphanumeric key codes implemented here are speciic to the German keyboard
layout. The following alphanumeric characters are used (some require the "Shift" key): a …
z, A … Z, 0 … 9, <space>, and these characters: ,.\+’<>/»$@%/();=:_?*
Function keys:
PC keyboard
Combics 2
PC keyboard
Combics 2
F1
key
F10
key
F2
key
F11
key
F3
key
F12
key
F4
key
key
F5
key
Return
key
F6
key
Pos 1
key
F7
key
Backspace
key
F8
key
ESC
key
F9
key
Combics 2 Series Complete Scales Models CAW2P, CAW2S
8 Data interfaces
EN-182