11 Error messages
A diferentiation is made between the following errors:
-
Dynamic errors are displayed with an error code (e.g. INF 01) for the duration of the
error.
-
Temporary errors are displayed for 2 seconds (e.g. INF 07)
-
Fatal errors are displayed permanently (e.g. ERR 101, can only be rectiied via reset)
Display
Cause
Remedy
No display seg-
ments
There is no operating voltage
Check electrical supply
-------
Key has no function in this state
Flashing
Battery defective or time adjusted
Set time
H
Weighing range has been exceeded
Unload the scale
L or ERR 54
Load plate is not positioned on the weighing
platform
Place the load plate onto the weighing
platform
ERR 101–104
Key stuck
Key pressed when switching on
Release key or
consult Minebea Intec customer service
ERR 320
Operating program memory defective
consult Minebea Intec customer service
ERR 335
Veriiable weighing platform does not match the
connected terminal
Connecting a suitable weighing platform
ERR 340
New EEPROM loaded (service)
Switch the scale of and on again; if
ERR 340 is permanently displayed, cont-
act Minebea Intec Customer Service
ERR 341
RAM lost data; battery empty
Leave device switched on for at least
10 hours
ERR 343
Data loss in the memory for transaction number
external Alibi memory
consult Minebea Intec customer service
INF 01
Data output does not it in the output format
Make correct setting in the setup menu
INF 02
Calibration condition was not complied with, e.g.
not tared or load plate loaded
Only adjust after zero display, unload sca-
le, tare with key
INF 03
The calibration process could not be completed
within a certain time.
Allow the device to warm up and then re-
calibrate the device
INF 06
Integrated calibration weight defective
consult Minebea Intec customer service
INF 07
The last used function is not allowed for veriied
scales
consult Minebea Intec customer service
INF 08
The load on the scale is too heavy to carry out ze-
roing
Please check whether the "power-on zero
range" (menu item [INIT.NUL] 1.12) was
complied with during the coniguration.
INF 09
No tare possible at gross zero
Zero the scales
Combics 2 Series Complete Scales Models CAW2P, CAW2S
11 Error messages
EN-196
Minebea Intec