— A.TARE
Automatic Tare
— OFF
disabled* (default setting)
— ON
first stable weight ≥ 2d is tared
— ON-ACC
Check: every stable weight within the accept
limits is tared. That means Auto-Tara isn’t done
only for one time!
Pieces as parameters (= pieces and weights are added up):
Printout
Description
N: 25
TOTAL: 148 g
124 pcs
MAX: 20 g
10 pcs
MIN: 4 g
2 pcs
Number of weighing done, here 25
Totalized value of weights
Totalized value of pieces
Maximum value as weight
Maximum value as number of pieces
Minimum value as weight
Minimum value as number of pieces
6.6
Auto-Tare
Application symbol
is on Display-Weight.
Weight
6.6.1
Menu Selection
OP.FUNC
6.6.2
Running Mode
Normal reaction:
-
First weight put on the pan is tared if it is ≥ 2d.
-
If the scale is unloaded (<2d) the Tara storage is cleared.
Check weighing applications:
-
First option [ON]:
First weight put on the pan is tared if it is ≥ 2d
-
Second option [ON-ACC]:
Only stable values within the accepted limit area is tared if load ≥ 2d. Here each
weight in the accepted limit area is tared.
This function is executed before automatic print and also before automatic Totalization!
Automatic Tare has higher priority as Printing or Totalization. First weight put on the pan
is tared. Second weight put on the pan is taken over into the totalization storage and is
also printed automatically when "Automatic print" and "Totalization Automatic" is
selected in menu.
Auto-Tare isn’t active if the menu is active or an initialization process is running.
Auto-Tara isn’t active if a Preset-Tare is set before or a product with Preset-Tare has been
loaded.
Compact Scale Puro® Count
6 Application Programs
EN-32
Minebea Intec