Checking additional weights then both limits must be a positive value.
Checking removed weights then both limits must be a negative value.
Using "Zero Check" then the high limit is a positive and the low limit is a negative value.
Checking exactly a weight the under and over limit must be the same.
6.4.2.2
Running Mode
Limit LEDs:
positive weight
< low limit
Yellow LED will light
positive weight
≥ low limit and ≤ high limit
Green LED will light
positive weight
> high limit
Red LED will light
negative weight
> low limit
Yellow LED will light
negative weight
≤ low limit and ≥ high limit
Green LED will light
negative weight
< high limit
Red LED will light
Using "Zero Check" the reference weight is put on the pan and then the scale is tared. The
reference weight is removed and then the weight to be checked is put on the pan.
4. Enter a numeric weight value via the key pad or a scanner which then is displayed on
Display-Count.
The input value isn’t blinking. Only the actual cursor position (an under line) is
blinking.
5.
Accept the input by pressing
(Yes) Button briefly or press the
(Back)
Button then the input is rejected.
Count
But if there is an error because e.g. high limit < low limit the Display-Count shows
[LIM.ERR] for about 2 seconds then go back to step 1.
The normal Counting mode is active again. Additional the Check LEDs are active
if the pan is loaded.
If the limit conditions are correct then Check Weighing is initialized.
6 Application Programs
Compact Scale Puro® Count
Minebea Intec
EN-25