Description
Data
Weight
Approx. 33 g
Certiicate
ProiBus Nutzerorganisation e.V. for HMS Industrial Net-
works AB
Certiicate no.: Z10931
Report: PN005
‑
1, 02/12/2007.
Note:
The ProiNet I/O card PR 1721/76 is supported by software version 2.24 and device serial
number 30363xxxxx or higher.
The IP address and subnet mask are set under [Fieldbus parameters] (see the
PR 5900 operating manual under [System setup] - [Fieldbus parameters] - [ProiNet I/
O] )
The XML ile is stored on the CD supplied with the device (ieldbus directory of the
respective device). The current ile is also available to download online:
http://www.minebea-intec.com
Note:
Fieldbus parameters
Recommendation for a Siemens S7, for example
Fieldbus slave setting:
Use DHCP [on] as per the default settings and activate the master as a DHCP server
(W [Allocate IP adr via IO controller]).
NOTICE
Slave – master device names
A unique device name must be assigned out of the master. This name is given highest
priority when establishing a connection.
When replacing devices or servicing, please note:
As well as the IP address, the device name must correspond to that of the
replacement device. Explicit assignment out of the master is required.
Example:
4 Device installation
Process Controller Maxxis 5 PR 5900
Minebea Intec
EN-113
Summary of Contents for Maxxis 5
Page 137: ......