NOTICE
For protection against dust and moisture during transport and installation, the
cable glands are itted with a polyethylene cover.
For full IP protection, operation with the dust protection cover itted is not permitted.
Remove the dust protection cover.
If a cable gland is not used, it must be sealed with a supplied locking pin.
NOTICE
Property damage is possible.
Regularly check the itted cable gland for tightness and re-tighten it, if necessary.
Cabling
In principle, the cables can be taken through all cable glands corresponding to the cable
diameters. The following igure shows a proposed topology.
11
12
6
7
8
9
10
5
4
3
2
1
No.
Description
1
WP B: 1 load cell or junction box
2
USB: barcode scanner/options M39/N29
3
Equipotential bonding conductor
4
Digital inputs and outputs: built-in/option
‑
1
5
WP A: 1 load cell or junction box
6
Power supply: 230 V AC or 24 V DC
7
Option
‑
1/ieldbus cards, except ProiBus
8
Option
‑
1/ProiBus card, when fully equipped
9
Ethernet socket (options M39/N29)
Process Controller Maxxis 5 PR 5900
4 Device installation
EN-30
Minebea Intec
Summary of Contents for Maxxis 5
Page 137: ......