6. Operating | 6.14 - Calibrating internal weighing point | 104
[LC Output at max. capacity]
Only with selection [All load cells the same]
Input: refer to the load cell certificate under "Output at max. capacity"
[LC output impedance]
Only with selection [All load cells the same]
Input: refer to the load cell certificate under "Output impedance"
[LC 1 to 8 Output at max. capacity]
Only with selection [Each load cell individual]
Input: refer to the load cell certificate under "Output at max. capacity"
[LC 1 to 8 output impedance]
Only with selection [Each load cell individual]
Input: refer to the load cell certificate under "Output impedance"
Note:
LC = Load cell
3. Save the adjustment.