With legal-for-trade applications, the CAL switch must be sealed in the closed position.
Note:
If the weighing electronics board has been changed after calibration or the device is not
calibrated the weight value display will show "E:BadDev" if the CAL switch is closed.
The position of the CAL switch is shown with VNC/WEB browser under
– [Show
status]:
[opened] = opened, no write protection.
[closed] = closed, write protection is active.
With the CAL switch in the 'closed' position, the weight is inversely shown in the display of
the device.
7.1.3.2
Factory settings (defaults)
Calibration data <default>
Calibration parameter <default>
Upper scale limit (Max) <3000>
<Kg>
Measurement time <160> ms
Scale interval <1>
Digital filter <off>
Dead load <0.000000> mV/V
Test mode <Absolute>
Span <1.000000> mV/V
W&M* <none>
Standstill time <0.5> s
Standstill range <1.00> d
Zero-setting range <50.00> d
Automatic zero tracking range <0.25> d
Scale interval for automatic zero tracking <0.25> d
Overload (range > Max) <9>d
Min <20> d
* Parameter [W&M] must be set to [none] before the calibration data is entered; see
Chapter
.
7 Commissioning
Transmitter in field housing PR 5230
EN-107
Minebea Intec