— Param
Configuration, "xBPI-Scale" selected: Assigned
to, Baudrate, Bits, Parity, Stopbits, see
Chapter
— Assign
Assign, "Pendeo Load Cells" selected: Search,
View, Calib, LC name, Service, see
Chapter
— Calib
Calibration of the weighing electronics
— New
Query window for new calibration: Reset Span
and dead load
Contin, Cancel
— Max
Input of the maximum load:
0.00001
<3000>
999999 <kg>, t, lb, oz, g, mg
— Scale interval
Input of the scale interval: <1>, 2, 5, 10, 20, 50
— Dead load at
<0.000000 mV/V> or [by load]
[by load]: 0.00001
999999 <kg>, t, lb, oz, g,
mg
— Max at
<1.000000 mV/V> or [by load]
[by load]: 0.00001
999999 <kg>, t, lb, oz, g,
mg
— Calibrated at
Display only
— Sensitivity (µV/d)
Display only
— Test
Determine test value
— Exit calibration
Save or discard changes due to new calibration.
— Modify
May be used only for minor changes (e.g.
changing the dead load, adapting mV/V values
for dead load and/or Max). Otherwise, always
use [New]!
— Param
Parameter settings
— Measuretime
Input of the measuring period: 5, 10, 20, 40, 80,
160, <320>, 640, 960, 1280, 1600 ms
— Digital filter
Selection of the digital filters: <off>, Bessel,
aperiod., Butterw., Tcheby.
— External supply
Selection: below or equal 8 V (≤8 V), <above 8
V> (>8 V)
— Fcut
Input of cut off frequency, only unless filter not
"off", 0.1–80.0 Hz
— Test mode
Selection for display of the deviation from the
test value: <Absolute>, relative
— W&M
Selection legal-for-trade mode: <none>, OIML
(impossible when [Range mode] "Multi-
interval" has been selected or Max has more
than 3 decimals), NSC, NTEP
— Standstill time
Input of the standstill period: 0.01 s
<0.50
s>
2.0 s (The range depends on the measuring
time.)
— Standstill range
Input of the standstill range: 0.00 d
<1.00
d>
10.00 d
— Tare timeout
Input of timeout when there is no standstill: 0.1
s
<2.5 s>
25 s
7.11.7.1
"Internal A" weighing point