[Measuretime]
Measuring time: The duration of a measurement can be selected.
Selection: 5 ms, 10 ms, 20 ms, 40 ms, 80 ms, 160 ms, <320 ms>, 640 ms, 960 ms,
1280 ms, 1600 ms.
[Digital filter]
Selection of the digital filter (filter characteristic): <off> (no Filter), Bessel, aperiod.
(aperiodic), Butterw. (Butterworth), Tcheby. (Tschebyscheff)
The following includes examples of interference signals for the different filter types:
Bessel filter
Aperiodic filter
Butterworth filter
Tschebyscheff filter
A digital filter can be switched on only with the measuring time set to <160 ms.
If no particularly frequent fluctuations are expected in ongoing operation, the following
settings are recommended:
[Measuretime]: <160 ms
[Digital filter]: aperiod.
[Fcut]: 2.00 Hz
[Fcut]
This line only is shown if the digital filter is switched on.
The smaller the cut-off frequency, the slower the measurement and the more stable the
measurement result.
Transmitter in field housing PR 5230
7 Commissioning
Minebea Intec
EN-142