Note:
If the W&M mode [OIML] has been selected, selection of the multi-interval scale [Multi-
interval] is not possible.
When selecting [Multi-interval], the following message is displayed:
Range mode multi interval
not allowed as long as
W&M is set to OIML
The weight display header includes the current interval range (R1, R2, or R3), Max, Min,
and d (or e with instruments used in legal metrology) (Example: multi-interval scale in
range 2):
The parameters [Range limit 1] and [Range limit 2] are the interval ranges.
As soon as the gross weight exceeds range 1, the next highest range with the next highest
scale interval becomes valid (1->2->5->10->20->50).
Note:
During calibration, the multi-interval function is always switched off.
Example:
Range mode: "Multi-interval"
Interval range 1: 0
1500 kg (when calibrating set scale interval: 1 kg)
Interval range 2: 1500
2900 kg (next highest scale interval: 2 kg)
1.
Choose "Range mode" from the
-[Weighing point]- [Calib]- [Param] menu.
2.
Select "Multi-interval" and confirm.