Card test:
The device display shows the fieldbus card status.
Under
- [HW-Slots] - [Slot 2] all inputs and outputs are displayed.
Procedure:
Note:
Further details can be found in the supplementary application manual "How to
"
(available upon request from technical.support.hh@minebea-intec.com).
Note:
See Chapter "Description of the I/O Area"
Example:
The gross weight should be read.
I/O size = 8 bytes, counted from byte 0–7
7.15.6
Network parameters
Under this menu item, you can define the network parameters for the network
connections (built-in LAN adapter).
Open the menu via
- [Network parameter].
[HW address]
This parameter cannot be changed because the fixed address is specified by the
instrument.
1.
Enter the IP address and network mask under
- [Fieldbus parameter] and
confirm.
2.
Register the file "sag_5230_Ethernetip.eds" using the "Hardware Installation Tool".
3.
Select and insert the instrument from the catalog into the I/O configuration.
7 Commissioning
Transmitter in field housing PR 5230
EN-193