Note:
VNC text, see also Chapter
LED
E:Arith
E:Invrs
E:Overl
E:NoSig
E:Sense
E:HardE/ E:NoCom
Flash. 1Hz
Flash. 1Hz
Altern. flash. 1Hz
Altern. flash. 1Hz
Flash. 1Hz
Flash. 1Hz
Altern. flash. 1Hz
Flash. 1Hz
Flash. 1Hz
Flash. 1Hz
Flash. 1Hz
Altern. flash. 1Hz
Altern. flash. 1Hz
Note:
With all other messages, the upper status indicator LED blinks.
16.3 Error messages for xBPI scales
Display
VNC text
Error and possible cause
Remote dis-
play
E:NoSig
Ext. meas. device error
No weight values can be read from scale:
-
Scale error
Error 3
E:Value
Value exceeds display
The weight value is not displayed:
-
Too many digits have been set.
Error 4
E:NoCom
No values from scale
No communication with xBPI scale:
-
Cable break.
-
Internal scale error.
-
The scale is not connected to the power sup-
ply.
Error 9
E:NoWgt
No weight data
No weight display:
-
Another weighing point was selected.
Press
to assign the new weighing point to
the device.
Error •
16 Error messages
Transmitter in field housing PR 5230
EN-287
Minebea Intec