Switch setting
PR 5230-1
Switch setting
PR 5230-2
Switch setting
PR 5230-3
Configuration
ON: S1, S2, S3, S5, S7
OFF: S4, S6, S8
ON: S1, S2, S3, S5, S7
OFF: S4, S6, S8
ON: S1, S2, S3, S4, S5, S7
OFF: S6, S8
- [Serial ports parameter]-
[
]- [Builtin RS485]
4.4.6.5
Connecting the remote display PR 5110
The PR 5110 remote display can be connected via the RS-485 interface.
Four-wire transmission, point to point connection, full duplex (simultaneous sending and
receiving) is possible with the remote display.
Switch settings PR 5230
PR 5110 switch settings
ON: S1, S2, S3, S6
OFF: S4, S5, S7, S8
ON: S1, S2, S3
OFF: S4, S5
Configuration PR 5230
PR 5110 configuration
- [Serial ports parameter] - [Remote display]
- [Builtin RS485]
[Param] - [Mode] - [single transmitter]
[
] - [oP 10] - [LInE] - [rS485]
[
] - [oP 12] - [tokEn] - [oFF]
[
] - [oP 13] - [SEndModE] - [SEnd]
[
] - [oP 14] - [WEIght] - [FolloW]
[
] - [oP 15] - [WPkEy] - [SElEct]
4 Device installation
Transmitter in field housing PR 5230
EN-49