5 Cable connections
5.1 General information
In order to use the junction boxes for load cells with 6-wire cables, jumpers J1 and J2 must
be opened; see Chapter
.
Cable entry must be from below.
Fit the wires and screen of the connection cable with wire end ferrules as per DIN 46228-1:
-
0.5 mm
2
for green, gray, black, and white
-
0.75 mm
2
for blue and red
This is not necessary for the load cell cables.
Connect the wires to the terminals according to the color coding .
Connect the screens on the other side of the connection cable with the equipotential
bonding terminal of the downstream device (see instrument manual) as described in
Chapters
and
Note:
If hum interference occurs, the cable screens should only be connected on one side.
Depending on the design of the cable junction box used, either the jumper J3 must be
removed or the cable screens must be disconnected from the terminal contacts
highlighted in yellow.
5.2 Connecting cables
To connect the junction box to the device, connection cable PR 6135/.. must be used;
max. length: 300 m.
5.3 Cable connections
Note:
All components are only shown schematically.
Color code
bk
=
black
bu
=
blue
gn
=
green
gy
=
gray
rd
=
red
wh
=
white
Cable Junction Box PR 6130/34Sa
5 Cable connections
EN-14
Minebea Intec
Summary of Contents for PR 6130/34Sa
Page 17: ...Connection example 5 Cable connections Cable Junction Box PR 6130 34Sa Minebea Intec EN 15 ...
Page 24: ...8 1 MEU17042 Cable Junction Box PR 6130 34Sa 8 Certificates EN 22 Minebea Intec ...
Page 25: ...8 Certificates Cable Junction Box PR 6130 34Sa Minebea Intec EN 23 ...
Page 26: ...Cable Junction Box PR 6130 34Sa 8 Certificates EN 24 Minebea Intec ...
Page 27: ...8 Certificates Cable Junction Box PR 6130 34Sa Minebea Intec EN 25 ...
Page 28: ...Cable Junction Box PR 6130 34Sa 8 Certificates EN 26 Minebea Intec ...
Page 29: ...8 Certificates Cable Junction Box PR 6130 34Sa Minebea Intec EN 27 ...