Continue for "dead load with empty scale":
Defining the span (Cd004)
Continue for "range with weights":
dEAdLo
appears on the display.
22. Press OK.
LoAd
(dead load with empty scale) appears on the display.
23. Press
to select the "mV/V" entry if necessary.
MVolt
appears on the display.
24. Press OK.
unLoAd
appears on the display.
25. Empty the scale.
26. Press OK.
A weight value for the dead load appears on the display.
27. Press OK.
00000.0 kg
appears on the display.
The dead load is set.
28. Press OK.
Cd 004
appears on the display.
29. Press OK.
SPAn
appears on the display.
30. Press OK.
LoAd
(range with weights) appears on the display.
31. Press
to select the "mV/V" entry if necessary.
MVolt
appears on the display.
32. Press OK.
33. Place the calibration weight on the scale.
A weight value that has not been calibrated yet appears on the display.
34. Press OK.
WEIGHt
appears on the display.
35. Press OK.
7 Commissioning
X3 Process Indicator PR 5410
EN-129
Minebea Intec