Input of timeout when there is no standstill: 0.1
s
<2.5 s>
25 s
— Zeroset range
Input of timeout period when there is no
standstill: 0.00 d
c
<50.00 d
c
>
10000.00 d
c
— Zerotrack
range
Input of the zerotrack range: 0.00 d
c
<0.25
d
c
>
10000.00 d
c
— Zerotrack step
Input of the zerotrack step: 0.00 d
c
<0.25
d
c
>
10.00 d
c
— Zerotrack time
Input zerotrack time: <0.0 s>
25 s
— Overload
Input of the weighing range above the
maximum load (Max) without error message:
0
<9 d
c
>
999999 d
c
— Min
Input of the minimum load: 0 d
c
<50
d
c
>
999999 d
c
— Range mode
Range selection: <Single range>, Multiple
range, Multi-interval
See also Chapters 7.14.15.2 and 7.14.15.3.
— Range limit 1
Input of the limit 1: In weight, unit same as Max,
transition from small to medium verification
interval
Only for [Multiple range] or [Multi-interval]!
— Range limit 2
Input of the limit 2: In weight, unit same as Max,
transition from small to medium verification
interval
Only for [Multiple range] or [Multi-interval]!
— LC name
Give each load cell a name.
— LC 1…n
load cell 1…n, e.g.: PR6224-xx
Input of max. 20 alphanumeric characters.
— Default
Settings are reset to factory settings.
— Service
Service function for load cells: Deactivate/
activate load cell.
— LC 1…n
load cell 1…n
Select the faulty load cell and reset ☑ to ☐.
Select the new (replaced) load cell and mark the
checkbox ☑.
— Accept
Accept:
After deactivation the simulation for the
deactivated load cell starts.
Once the replaced load cell has been activated
the search process starts.
— Assign (CAL-switch is closed)
— View
Item number, serial number and current weight
of connected load cells are displayed.
— Info
The data for the selected Loading cell are
displayed.
— by name
Change the display from [ID] (LC 1…n + serial
number) to [by name] of the load cells; only
possible if a load cell name has been assigned.
— LC name
Change the display from [ID] (LC 1…n + serial
number) to [by name] of the load cells; only
possible if a load cell name has been assigned.
X3 Process Indicator PR 5410
7 Commissioning
Minebea Intec
EN-156