Butterworth filter
Tschebyscheff filter
A digital filter can be switched on only with the measuring time set to <160 ms.
If no particularly frequent fluctuations are expected in ongoing operation, the following
settings are recommended:
[Measuretime]: <160 ms
[Digital filter]: aperiod.
[Fcut]: 2.00 Hz
[Fcut]
This line only is shown if the digital filter is switched on.
The smaller the cut-off frequency, the slower the measurement and the more stable the
measurement result.
The cut-off frequency can be specified for the low pass filter.
Valid range: 0.1
2.5 Hz.
The available options depend on the measuring time.
[External supply]
When the load cells are connected to an external supply, it is possible to switch to ≤8 V, to
tune Sense voltage monitoring to the lower supply voltage.
Selection: below or equal 8 V (≤8 V), <above 8 V> (>8 V)
[Test mode]
With [absolute], the test value is determined when the test is called.
With [relative], the deviation from the initially stored test value is displayed; see
Chapter
[W&M]
See Chapter
If [OIML] was selected, the device needs to warm up for 30 seconds.
[Standstill time]
The parameters [Standstill time] and [Standstill range]are used to define the stability of
the scale (stable balance).
The input for the parameter [Standstill time] is expressed in seconds.
Valid range: 0.00
2 s
If the tome is set to "0" there is no check. The standstill time must not be less than the
measuring time.
X3 Process Indicator PR 5410
7 Commissioning
Minebea Intec
EN-174