[Standstill range]
As long as the weight fluctuations remain within this range, the device is determined to
be stable.
The input for the parameter [Standstill range] is expressed in "d."
Valid range: 0.01
10.00 dc.
[Tare timeout]
Timeout for a tare/zeroset command that cannot be executed (e.g. due to mechanical
instability of the scale, incorrect filter setting, resolution too high, standstill condition too
strict).
The input is done as seconds.
Valid range: 0.0
<2.5>
25 s.
At 0.0 s taring is only carried out when the scale is already stable.
[Zeroset range]
Define a ±range around the zero point determined by the dead load during calibration;
within this range
-
the displayed gross weight can be set to zero by pressing the zero-setting key (or by
a corresponding external command), and
-
automatic zero tracking is active.
Setting range: 0.00
10000.00 d
For use in legal metrology a value ≤2% of the Max must be entered, example: 60 d for
3000 e of class III.
[Zerotrack range]
Range within which automatic zero tracking compensates deviations.
Setting range: 0.25
10000.00 d
In "legal-for-trade" mode a value of 0.25 d has to be entered.
[Zerotrack step]
If a weight change exceeds the adjusted value, automatic tracking does not function any
more.
Setting range for automatic tracking increments: 0.25
10 d
In "legal-for-trade" mode a value of 0.25 d has to be entered.
[Zerotrack time]
Time interval for automatic zero tracking.
Setting range: 0.1
25 s
At 0.0 s the zero tracking is switched off.
For use in legal metrology a value of 1 s must be entered.
[Overload]
Weighing range above the maximum load (Max) without error message.
Setting range: 0
9999999 d
For use in legal metrology a value of max. 9 d = e must be set.
[Minimum weight]
Minimum weight at which a print command can be triggered.
Setting range: 0
9999999 d
For use in legal metrology a value of at least 20 d must be set.
7 Commissioning
X3 Process Indicator PR 5410
EN-205