[Range mode]
Selection: <Single range>, Multiple range, Multi-interval
For scale range selection, see Chapter
and
Press
to exit the menu and to save the settings.
7.16.13
Subsequent dead load correction
If the hopper/platform weight changes by an amount that is higher than the zero-setting
range; e.g. due to dead load reduction, dead load increase, or mechanical changes, the
functions for automatic zero tracking and manual zero setting no longer work.
To view the range which is already utilized by zero tracking or zero setting, in [Calibration]
press the
key; this also activates 10-fold increased resolution of the weight value.
Press
again to return to the previous state.
Note:
The scale must not be loaded!
If the entire zero-setting range is already utilized, you can still correct the dead load
subsequently without affecting other calibration data/parameters. To do this open
calibration via
-[Weighing point]- [Weighing point A]- [Assign]- [Calib]- [Modify] and
determine the dead load with [Dead load] using the [by load] option (see Chapter
7.16.14
Displaying weighing point serial number
After searching via
-[Weighing point]- [Weighing point A]- [Assign]- [Search] the
corresponding weighing point serial number is displayed.
7.17 General parameter settings
The parameter settings which are not related to the weighing electronics are divided into
several ranges.
-
Serial interfaces [Serial ports parameter]
-
Date and Time [Date & Time]
X3 Process Indicator PR 5410
7 Commissioning
Minebea Intec