9.4
Resetting the device to the factory settings
Note:
Reset to the factory settings is possible only, when the CAL switch is open. The IP
address and the Hostname remain unaffected.
1.
Press
.
SEtuP
appears on the display.
2.
Press
.
bIoS
appears on the display.
3.
Press OK.
FLASH
follows on the display.
bIoS...
appears on the display.
4. Press
2×.
dEL.EAr.
appears on the display.
5.
Press OK.
ErAS.E?
and then
no
appear on the display.
6. Press
.
YES
appears on the display.
7.
Press OK.
donE
appears on the display once the process is complete.
8. Press
to restart the device.
1.
Click on
.
2.
Click on
.
The following prompt window appears:
X3 Process Indicator PR 5410
9 Extended functions
Minebea Intec
EN-280