9.5.2.1
Operations at the device
WARNING
Working on the device while it is switched on may have life-threatening
consequences.
Product operation must be performed by trained and qualified personnel who are
aware of and able to deal with the related hazards and take suitable measures for
self-protection.
1.
Press
.
SEtuP
appears on the display.
2.
Press
.
bIoS
appears on the display.
3.
Press OK.
FLASH
follows on the display.
bIoS...
appears on the display.
4. Press OK.
EtHEr
appears on the display.
5.
Press OK.
The device waits for address assignment by the server.
172.024.
The more significant part of the IP address is displayed for approx.
2 seconds.
dHCP...
appears on the display.
6. Make sure to note down the digit sequence, because it must be entered into the
"FlashIt!32" window subsequently.
021.081.
The less significant part of the IP address is displayed for approx. 2
seconds.
7.
Make sure to note down the digit sequence, because it must be entered into the
"FlashIt!32" window subsequently.
FLASH?
appears on the display.
X3 Process Indicator PR 5410
9 Extended functions
Minebea Intec
EN-284