4.6.3.2.3
Connecting a PR 5110 remote display
The PR 5110 remote display can be connected via the RS-485 interface.
Four-wire transmission, point-to-point connection, full duplex (simultaneous sending and
receiving) is possible with the remote display.
Example:
Switch settings PR 5410
PR 5110 switch settings
ON: S1, S3, S4, S5
OFF: S2
ON: S1, S2, S3
OFF: S4, S5
Configuration PR 5410
PR 5110 configuration
- [Serial ports parameters]- [Remote dis-
play]- [Slot 1/2 RS-485]
[Param]- [Mode]- [single transmitter]
[
]- [oP 10]- [LInE]- [rS485]
[
]- [oP 12]- [tokEn]- [oFF]
[
]- [oP 13]- [SEndModE]- [SEnd]
[
]- [oP 14]- [WEIght]- [FolloW]
[
]- [oP 15]- [WPkEy]- [SElEct]
The following operations are possible from the connected remote display:
-
Display current weight mode
-
Set tare
-
Reset tare
-
Set zero
-
Start printout
X3 Process Indicator PR 5410
4 Device installation
Minebea Intec
EN-66