Changing the dead load (Cd 003)
Saving and exiting
7.5.4
Displaying the calibration data of the internal WP using the front-panel keys
5.
Press
to change the parameter number ("000" flashes).
6. Press
to select the parameter number.
7.
Press OK.
dEAdLo
appears on the display.
8. Press OK.
LoAd
(dead load with empty scale) appears on the display.
9. Press OK.
unLoAd
appears on the display.
10. Empty the scale.
11. Press OK.
A weight value for the dead load appears on the display.
12. Press OK.
00000.0 kg
appears on the display.
The dead load is set.
13. Press OK.
Cd 004
appears on the display.
14. Press "Exit."
SAVE
appears on the display.
15. Press OK.
YES
appears on the display.
16. Press OK.
SAVE ...
appears on the display during the save process.
17. Press "Exit" to exit the Setup menu.
18. If no other calibration data or parameters are going to be set, the CAL switch must be
closed now; see Chapter
.
1.
Press
.
SEtuP
appears on the display.
7 Commissioning