To use the empty scale/hopper as dead load (normal case):
Note:
If the mV/V value of the dead load was calculated, or if it is known from the previous
calibration, the value can be overwritten by pressing [by mV/V].
Error messages, calibration see Chapter
.
7.14.7
Calibrating with weight
Note:
If a linearization was carried out (see Chapter
), after selection of the line [Max at]
the following tip is displayed:
Cannot be changed here
while linearization is active.
Changes cannot be made while linearization is switched on.
Only deleting of the linearization points deactivates the linearization mode!
1.
Clear the scale/hopper.
2.
Press the [by load] softkey.
3.
Confirm entries with
or
.
The verification is displayed by "Setting dead load…".
7 Commissioning